ORIGINAL ARTICLE
Year : 2022  |  Volume : 13  |  Issue : 1  |  Page : 55-61  

Comparison of in vivo kinematic and radiological parameters of three cervical disc prostheses


NeuroSpine Surgery Research Group, Prince of Wales Private Hospital, Randwick; Faculty of Medicine, University of New South Wales, Sydney, NSW, Australia

Date of Submission28-Jun-2021
Date of Acceptance02-Oct-2021
Date of Web Publication9-Mar-2022

Correspondence Address:
Henry Lin
NeuroSpine Surgery Research Group, Prince of Wales Private Hospital, Randwick 2031, NSW
Australia
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/jcvjs.jcvjs_92_21

Rights and Permissions
   Abstract 


Introduction: Cervical total disc replacement (CTDR) is an alternative to anterior cervical discectomy and fusion for select patients that may preserve range of motion and reduce adjacent segment disease. Various CTDR prostheses are available; however, comparative data are limited. This study aimed to compare the short-term kinematic and radiological parameters of the M6-C, Mobi-C, and the CP-ESP prostheses.
Methods: This retrospective cohort study included patients treated with CTDR between March 2005 and October 2020 at a single institution. Patients were included if their follow-up assessment included lateral erect and flexion/extension radiographs. The primary outcome assessed at 3-months postoperatively was range of motion, measured by the difference in functional spinal unit angle between flexion and extension.
Results: A total of 131 CTDR levels (120 patients, 46.2 ± 10.1 years, 57% male) were included. Prostheses implanted included the M6-C (n = 52), Mobi-C (n = 54), and CP-ESP (n = 25). Range of motion varied significantly (8.2° ± 4.4° vs. 10.9° ± 4.7° vs. 6.1° ± 2.7°, P < 0.001). On post hoc analysis, the Mobi-C prosthesis demonstrated a significantly greater range of motion than either the M6-C prosthesis (P = 0.003) or CP-ESP (P < 0.001).
Conclusion: Although the optimal range of motion for CTDR has not been established, short-term differences in the range of motion may guide the selection of CTDR prosthesis. Further studies with longer follow-up and consideration of clinical outcome measures are necessary.

Keywords: Cervical total disc replacement, CP-ESP, disc arthroplasty, M6-C, Mobi-C, total disc replacement


How to cite this article:
Chang N, Mobbs R, Hui N, Lin H. Comparison of in vivo kinematic and radiological parameters of three cervical disc prostheses. J Craniovert Jun Spine 2022;13:55-61

How to cite this URL:
Chang N, Mobbs R, Hui N, Lin H. Comparison of in vivo kinematic and radiological parameters of three cervical disc prostheses. J Craniovert Jun Spine [serial online] 2022 [cited 2022 Dec 6];13:55-61. Available from: https://www.jcvjs.com/text.asp?2022/13/1/55/339305




   Introduction Top


Cervical total disc replacement (CTDR) has emerged as an effective alternative to anterior cervical discectomy and fusion (ACDF) for some patients with cervical degenerative disc disease resulting in radiculopathy or myelopathy.[1] Although ACDF has traditionally been the procedure of choice, the resultant loss of mobility may result in compensatory changes in adjacent segments and accelerated degenerative pathology, so-called adjacent segment disease (ASD).[2],[3] In contrast, CTDR aims to relieve the symptoms and preserve physiological mobility of the cervical spine.[4],[5] Biomechanical evidence suggests that CTDR may result in decreased adjacent level stress,[6],[7],[8] and clinical studies have demonstrated decreased radiographic evidence of ASD compared to ACDF.[9] The relationship between radiographic and clinical ASD is incompletely understood,[10] but there is also growing evidence that CTDR may also decrease the incidence of further surgery at adjacent levels.[11],[12],[13]

There are numerous cervical arthroplasty prostheses available, utilizing a variety of different design principles.[1] The three commonly used prostheses in Australia include the Mobi-C (LDR Spine USA, Austrin, TX, USA), M6-C (Spinal Kinetics, Sunnyvale, CA, USA) and the recently developed CP-ESP (FH Orthopedics, Mulhouse, France). Each of these devices has been investigated individually, but there is a lack of comparative evidence.[13],[14],[15],[16] This study aims to define and compare the short-term in vivo biomechanical and radiological properties of these three CTDR prostheses.


   Materials and Methods Top


A retrospective cohort study was conducted on patients derived from a prospectively maintained database. All patients were treated at a tertiary hospital by a single senior spine surgeon. Ethics approval was received from the Institutional Review Board of the authors' affiliated institution. All patients provided consent for the release of their medical data.

Population

This study included consecutive adult patients who underwent CTDR between March 2005 and August 2020 who satisfied the exclusion criteria [Table 1] and had adequate 3-month follow-up including lateral flexion/extension radiographs of the cervical spine [Figure 1]. This study included no restrictions on the number of cervical levels treated or the existence of previous surgery at other cervical spine levels. The indication for surgery was symptomatic radiculopathy and/or myelopathy secondary to cervical degenerative disc disease, which did not respond to conservative treatments for at least 3 months. Three types of CTDR procedure were included in this study: Single-level CTDR, multi-level CTDR, and hybrid surgery (combined CTDR and ACD). Each CTDR operative level was treated as an independent observation.
Table 1: Exclusion criteria

Click here to view
Figure 1: Lateral radiographs demonstrating flexion (left) and extension (right) for a range of patients with a C5/6 single level total disc replacement who received (a) M6-C prosthesis, (b) Mobi-C prosthesis, and (c) CP-ESP prosthesis

Click here to view


Radiographic assessment

Biomechanical measurements were recorded based on lateral flexion/extension views of standing radiographs, using Cobb angle measurement tools inbuilt into radiological analysis software (InteleRad, Canada). Range of motion was assessed by measuring the difference in functional spinal unit (FSU) angle in flexion and extension. The FSU angle was defined by the angle formed between lines subtended from the superior endplate of the rostral vertebral body and the inferior endplate of the caudal vertebral body [Figure 2]. A lordotic angle was defined to be a positive value. The anterior and posterior disc height of each operated level was also measured on an erect lateral radiograph and defined as the distance between the inferior endplate of the cranial vertebral body and the superior endplate of the caudal vertebral body at the anterior and posterior limits of the disc space, respectively [Figure 2].
Figure 2: Radiological assessment methods: (a) Functional spinal unit angle, measured on flexion/extension lateral radiographs extension angle using Cobb angle measured tool and defined by the angle formed by lines subtended from the superior endplate of the rostral vertebral body and inferior endplate of the caudal vertebral body; (b) anterior and posterior disc heights, measured on an erect lateral radiograph and defined as the distance between the inferior endplate of the cranial vertebral body and the superior endplate of the caudal vertebral body at the anterior and posterior limit of the disc space, respectively

Click here to view


Surgical approach

A standard right-sided anterior approach to the cervical spine was utilized in all cases, as previously described.[17] After dissection, a Caspar distractor (A-Spine ASIA, Taiwan) was used to distract the disc space at the operative level and a discectomy was performed. The endplates were prepared for prosthesis insertion. A template prosthesis was used to confirm sizing, and then, the appropriate prosthesis was inserted with intraoperative radiography utilized throughout the procedure. The types of implants used during the trial period included the Mobi-C (LDR Spine USA, Austrin, TX, USA), M6-C (Spinal Kinetics, Sunnyvale, CA, USA), and the CP-ESP (FH Orthopedics, Mulhouse, France). The implant used was based upon the preference of the patient and surgeon during a preoperative discussion, with consideration to the range of implants available at the time and individual patient factors. The patient received a soft collar and routine anti-inflammatory medication for 2 weeks postoperatively.

Statistical assessment

Statistical analysis was performed using the SPSS software version 26 (IBM; Armonk, New York, USA). P < 0.05 was considered statistically significant. Parametric continuous data were expressed as mean ± standard deviation and categorical data were expressed as percentages. Parametric continuous variables were compared across the three prostheses using Welch's or standard one-way analyses of variance, depending upon whether the assumption of homogeneity of variances was violated or not. Post hoc testing was performed using either the GamesHowell test or the TukeyKramer test, depending upon whether the assumption of homogeneity of variances was violated or not. The categorical variables were compared across the three prostheses using the Chi-squared test or Fisher's exact testing as appropriate.


   Results Top


Characteristics of study population

A total of 131 CTDR levels were included in this study (120 patients, 46.2 ± 10.1 years, 57% male). Baseline characteristics are summarized in [Table 2]. Prostheses implanted included M6-C (n = 52), Mobi-C (n = 54), and CP-ESP (n = 25). Prostheses were implanted mostly in single-level CTDR procedures (56%), and less commonly in multi-level CTDR procedures (12%) or hybrid procedures (31%). All multi-level CTDR procedures included only two spinal levels while hybrid procedures included both two-level (83%) and three-level (17%) procedures. The most common spinal segments for CTDR prostheses in this study were C5/6 (52%) and C6/7 (26%).
Table 2: Characteristics of the study population

Click here to view


Range of motion

The range of motion for each implanted prosthesis at 3-month follow-up is summarized in [Table 3]. Across the three prostheses tested: M6-C, Mobi-C, and CP-ESP, range of motion varied significantly (8.2° ± 4.4° vs. 10.9° ± 4.7° vs. 6.1° ± 2.7°, P < 0.001). Post hoc analysis with the GamesHowell test revealed a significantly greater range of motion for the Mobi-C prosthesis compared to either the M6-C (P = 0.003) or the CP-ESP (P < 0.001). Range of motion data for the two most common surgical levels, C5/6 and C6/7, are also presented in [Table 3]. The comparison of prostheses implanted at C5/6 revealed significant variation between the three prostheses (P < 0.001) and a significantly greater range of motion for the Mobi-C prosthesis compared to the CP-ESP on post hoc testing with the TukeyKramer test (P < 0.001). Other pairwise comparisons at C5/6 were not significant. Range of motion at C6/7 did not vary significantly across prostheses (P > 0.05).
Table 3: Range of motion for cervical total disc replacement prostheses at 3 months follow-up

Click here to view


Disc height

The anterior and posterior disc heights achieved at 3-month follow-up are summarized in [Table 4]. Across the three prostheses tested: M6-C, Mobi-C, and CP-ESP, significant variation was detected in both anterior disc height (9.5 ± 1.4 mm vs. 8.5 ± 1.3 mm vs. 8.4 ± 0.9, P = 0.002) and posterior disc height (6.1 ± 1.1 mm vs. 5.4 ± 1.4 mm vs. 4.8 ± 1.6 mm P = 0.001). Post hoc comparisons with the GamesHowell test revealed that the M6-C prosthesis produced significantly greater anterior disc height at 3 months than the Mobi-C (P = 0.002) or CP-ESP prostheses (P = 0.007). Similarly, post hoc testing the TukeyKramer test revealed that the M6-C produced greater posterior disc height at 3 months than the Mobi-C (P = 0.023) or CP-ESP (P < 0.001).
Table 4: Anterior and posterior disc height for cervical total disc replacement prostheses at 3 months follow-up

Click here to view



   Discussion Top


This study assessed the short-term kinematic and radiological outcomes following CTDR with three commonly used prostheses: The M6-C, Mobi-C, and CP-ESP prostheses. This is the first study to provide comparative data between the CP-ESP and other prostheses. This study found significant variation between prostheses with respect to the range of motion at 3 months, with the Mobi-C demonstrating a greater ROM than either the M6-C or CP-ESP.

It is hypothesized that the preservation of physiological ROM following CTDR may reduce the risk of ASD compared to ACDF.[1] There is a growing body of evidence in support of this hypothesis, with previous biomechanical studies demonstrating reduced adjacent level intradiscal pressure and facet joint forces at adjacent levels after CTDR compared to ACDF[6],[7],[8] This has also been supported by clinical evidence, with most randomized controlled trials reporting significantly less ASD after CTDR compared to ACDF, in a recent systematic review and meta-analysis.[18] However, the relationship between postoperative range of motion at the operated segment and the clinical outcome has not been clearly established. This study found that the Mobi-C prosthesis produced a greater range of motion than the other studied prostheses. This finding aligns with previously published data. One previous comparative study of the M6-C and Mobi-C prostheses found that the Mobi-C resulted in significantly greater extension compared to the M6-C.[19] Similarly, a biomechanical study in 16 fresh-frozen human cervical spine specimens reported that implantation of a C5/6 Mobi-C prosthesis resulted in significantly greater segmental ROM compared to preoperative values (16.8° ± 3.7° vs. 13.8° ± 4.2°, P < 0.05), whereas the M6-C prosthesis did not result in a change from preoperative values.[20]

For healthy young adults, the normative mean ROM for a subaxial cervical motion segment varies between 10.2° and 19.7°.[21],[22],[23] According to previous normative studies, the mean ROM was observed to be between 14.6°–19.7° at C5/6, and 12.3°–15.8° at C6/7.[21],[22],[23] In contrast to these normative studies, the ROM for each prosthesis in this study appear to be lower, with mean ROM varying between 5.9°–11.7° at C5/6, and 6.2°–10.7° at C6/7. However, comparison with historical controls is difficult as differences in ROM could be accounted for by many factors, such as age,[24],[25] joint pathologies,[26] postural abnormalities,[24] and gender[25] as well as inter-rater variability in ROM measurement.

While the optimal ROM following CTDR is not yet clear, supra-physiological mobility [Figure 3] may worsen clinical outcome and increase the risk of ASD or facet joint degenerative changes. It has been postulated that excessive prothesis mobility may put increased demand on the spinal musculature to stabilize the spine, increasing strain on spinal components and contributing to pain.[27] It is possible that the ROM results of this study may inform surgeon or patient preference for a particular implant, but further studies with longer-term follow-up and consideration of clinical outcomes are required to determine the best available prosthesis. The limited published comparative data did not find a significant difference in clinical outcomes measures between the Mobi-C and M6-C prostheses.[17]
Figure 3: Lateral radiographs demonstrating hypermobility in flexion (left) and extension (right) at the operated level in patients who received (a) M6-C prosthesis and (b) Mobi-C prosthesis. No cases of hypermobility of the CP-ESP prosthesis were identified during the trial period

Click here to view


The prostheses used in this current study vary in their design principles and materials and this may have contributed to the ROM data observed. Both the M6-C and the CP-ESP employ endplate sandwiched viscoelastic designs which are able to provide six degrees of freedom and combined motions due to deformation of their viscoelastic components.[28] The M6-C is a mobile bumper viscoelastic device, featuring a compressible polycarbonate urethane polymer that mimics the nucleus pulposus and a polyethylene weave annulus that is intended to prevent tissue ingrowth and contain wear debris and is welded to titanium endplates.[29] The CP-ESP is a monoblock viscoelastic device featuring a polycarbonate urethane core fixed to titanium endplates.[15] There are theoretical advantages and disadvantages to each design: The M6-C has no stress peak at the interface of the core and endplate as these components are unbonded, while the CP-ESP, which has no motion at the core-endplate interface has a corresponding reduced risk of wear debris.[28] In contrast, the Mobi-C is a “ball and socket” prosthesis with a mobile core of polyethylene articulating with superior and inferior cobalt chromium molybdenum alloy endplates.[20],[30] The mobile core has a spherical articulation with the superior endplate and a planar articulation with the inferior endplate. The contrast in design between the viscoelastic prostheses (M6-C and CP-ESP), which provide increasing resistance to motion at the extremes of ROM, and the “ball and socket” prosthesis (Mobi-C) may partially explain the observed difference in range of motion.

Similarly, the variation in anterior and posterior disc height observed in this study may also be related to the design of the individual prostheses and the surgical technique necessary for their implantation. The Mobi-C and CP-ESP implants have an anatomical shape, with a flat inferior endplate and a domed superior endplate, and therefore, minimal endplate preparation is necessary for their implantation. The M6-C prosthesis has a wedge-shaped design with flat inferior and superior endplates and significant endplate preparation is necessary to provide space for implantation. In addition, the smallest height available for the M6-C prosthesis is 6 mm, while the height of a natural disc may vary between 3.5 and 6.1 mm.[31],[32],[33] As such, the height of the prosthesis and the need for significant tissue removal during endplate preparation likely contributed to the greater anterior and posterior disc height observed for the M6-C prosthesis. It has been postulated that excessive anterior/posterior disc height (over-distraction) may result in worsened clinical outcomes through stretching of facet joints.[28],[34] However, previous comparative studies of the M6-C and Mobi-C prostheses did not report significantly different clinical outcomes.[17]

Data comparing different CTDR prostheses is limited and does not show significant clinical differences between prostheses.[17],[19] However, each of the prostheses included in this study has demonstrated clinical efficacy in previous trials. The M6-C demonstrated positive clinical outcome results and no serious adverse effects in an early study of 36 one-level CTDR patients.[16] More recently, the M6-C has received Food and Drug Administration (FDA) approval for single-level CTDR based upon its positive clinical and safety profile in a US FDA Investigational Device Exemption (FDA IDE) trial that assessed 160 patients who received M6-C prostheses against an ACDF control group.[35] A postapproval study is ongoing. The Mobi-C prosthesis has demonstrated clinical efficacy for both single-level and two-level CTDR in multiple FDA IDE trials.[11],[13],[36],[37] In addition, a prospective uncontrolled study conducted in France reported favorable outcomes for both single-level and multi-level CTDR.[38] The CP-ESP has been investigated in one previous study with a biomechanical component and a clinical component of 71 implanted protheses with 12 months follow-up.[15] This study concluded that the CP-ESP demonstrated encouraging clinical results but recognized that further long-term clinical evaluation was required.

The present study has several limitations that require consideration. Only ROM and anterior and posterior disc height were measured. There are several kinematic parameters not addressed by the current study, including centers of rotation, lateral flexion and vertebral body translation. In addition, clinical data such as Visual Analog Scale scores or other patient-reported outcomes were not assessed. Further studies are necessary to more complete compare the kinematic and clinical properties of these prostheses. In addition, the short follow-up period another limitation as ROM and disc height has been demonstrated to change over time.[9],[15] Finally, the single surgeon single-institution study design and relatively young cohort may influence the external validity of these results.


   Conclusion Top


At 3 months follow-up, significant variation was noted in the range of motion across the three prostheses tested. The Mobi-C prosthesis demonstrated a significantly greater range of motion than the M6-C or CP-ESP. Short-term differences in ROM may guide the selection of CTDR protheses for various age groups and pathologies, although further research is required to determine the optimal ROM after CTDR.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
   References Top

1.
Nunley PD, Coric D, Frank KA, Stone MB. Cervical disc arthroplasty: Current evidence and real-world application. Neurosurgery 2018;83:1087-106.  Back to cited text no. 1
    
2.
Goffin J, Geusens E, Vantomme N, Quintens E, Waerzeggers Y, Depreitere B, et al. Long-term follow-up after interbody fusion of the cervical spine. J Spinal Disord Tech 2004;17:79-85.  Back to cited text no. 2
    
3.
Baba H, Furusawa N, Imura S, Kawahara N, Tsuchiya H, Tomita K. Late radiographic findings after anterior cervical fusion for spondylotic myeloradiculopathy. Spine (Phila Pa 1976) 1993;18:2167-73.  Back to cited text no. 3
    
4.
Park DK, Lin EL, Phillips FM. Index and adjacent level kinematics after cervical disc replacement and anterior fusion: In vivo quantitative radiographic analysis. Spine (Phila Pa 1976) 2011;36:721-30.  Back to cited text no. 4
    
5.
Wigfield C, Gill S, Nelson R, Langdon I, Metcalf N, Robertson J. Influence of an artificial cervical joint compared with fusion on adjacent-level motion in the treatment of degenerative cervical disc disease. J Neurosurg 2002;96 Suppl 1:17-21.  Back to cited text no. 5
    
6.
Chang UK, Kim DH, Lee MC, Willenberg R, Kim SH, Lim J. Changes in adjacent-level disc pressure and facet joint force after cervical arthroplasty compared with cervical discectomy and fusion. J Neurosurg Spine 2007;7:33-9.  Back to cited text no. 6
    
7.
Dmitriev AE, Cunningham BW, Hu N, Sell G, Vigna F, Mcafee PC. Adjacent level intradiscal pressure and segmental kinematics following a cervical total disc arthroplasty: An in vitro human cadaveric model. Spine (Phila Pa 1976) 2005;30:1165-72.  Back to cited text no. 7
    
8.
Eck JC, Humphreys SC, Lim TH, Jeong ST, Kim JG, Hodges SD, et al. Biomechanical study on the effect of cervical spine fusion on adjacent-level intradiscal pressure and segmental motion. Spine (Phila Pa 1976) 2002;27:2431-4.  Back to cited text no. 8
    
9.
Coric D, Cassis J, Carew JD, Boltes MO. Prospective study of cervical arthroplasty in 98 patients involved in 1 of 3 separate investigational device exemption studies from a single investigational site with a minimum 2-year follow-up. Clinical article. J Neurosurg Spine 2010;13:715-21.  Back to cited text no. 9
    
10.
Nunley PD, Jawahar A, Cavanaugh DA, Gordon CR, Kerr EJ 3rd, Utter PA. Symptomatic adjacent segment disease after cervical total disc replacement: Re-examining the clinical and radiological evidence with established criteria. Spine J 2013;13:5-12.  Back to cited text no. 10
    
11.
Radcliff K, Coric D, Albert T. Five-year clinical results of cervical total disc replacement compared with anterior discectomy and fusion for treatment of 2-level symptomatic degenerative disc disease: A prospective, randomized, controlled, multicenter investigational device exemption clinical trial. J Neurosurg Spine 2016;25:213-24.  Back to cited text no. 11
    
12.
Jackson RJ, Davis RJ, Hoffman GA, Bae HW, Hisey MS, Kim KD, et al. Subsequent surgery rates after cervical total disc replacement using a Mobi-C cervical disc prosthesis versus anterior cervical discectomy and fusion: A prospective randomized clinical trial with 5-year follow-up. J Neurosurg Spine 2016;24:734-45.  Back to cited text no. 12
    
13.
Hisey MS, Zigler JE, Jackson R, Nunley PD, Bae HW, Kim KD, et al. Prospective, randomized comparison of one-level Mobi-c cervical total disc replacement vs. anterior cervical discectomy and fusion: Results at 5-year follow-up. Int J Spine Surg 2016;10:10.  Back to cited text no. 13
    
14.
Lauryssen C, Coric D, Dimmig T, Musante D, Ohnmeiss DD, Stubbs HA. Cervical total disc replacement using a novel compressible prosthesis: Results from a prospective food and drug administration-regulated feasibility study with 24-month follow-up. Int J Spine Surg 2012;6:71-7.  Back to cited text no. 14
    
15.
Lazennec JY, Aaron A, Ricart O, Rakover JP. The innovative viscoelastic CP ESP cervical disk prosthesis with six degrees of freedom: Biomechanical concepts, development program and preliminary clinical experience. Eur J Orthop Surg Traumatol 2016;26:9-19.  Back to cited text no. 15
    
16.
Thomas S, Willems K, Van Den Daelen L, Linden P, Ciocci Mc, Bocher P. The M6-C cervical disk prosthesis: First clinical experience in 33 patients. Clin Spine Surg 2016;29:E182-7.  Back to cited text no. 16
    
17.
Hui N, Phan K, Kerferd J, Lee M, Mobbs RJ. Comparison of M6-C and Mobi-C cervical total disc replacement for cervical degenerative disc disease in adults. J Spine Surg 2019;5:393-403.  Back to cited text no. 17
    
18.
Findlay C, Ayis S, Demetriades AK. Total disc replacement versus anterior cervical discectomy and fusion: A systematic review with meta-analysis of data from a total of 3160 patients across 14 randomized controlled trials with both short- and medium- to long-term outcomes. Bone Joint J 2018;100-b: 991-1001.  Back to cited text no. 18
    
19.
Pham M, Phan K, Teng I, Mobbs RJ. Comparative study between M6-C and Mobi-C cervical artificial disc replacement: Biomechanical outcomes and comparison with normative data. Orthop Surg 2018;10:84-8.  Back to cited text no. 19
    
20.
Patwardhan AG, Havey RM. Prosthesis design influences segmental contribution to total cervical motion after cervical disc arthroplasty. Eur Spine J 2020;29:2713-21.  Back to cited text no. 20
    
21.
Anderst WJ, Lee JY, Donaldson WF 3rd, Kang JD. Six-degrees-of-freedom cervical spine range of motion during dynamic flexion-extension after single-level anterior arthrodesis: Comparison with asymptomatic control subjects. J Bone Joint Surg Am 2013;95:497-506.  Back to cited text no. 21
    
22.
Anderst WJ, Donaldson WF 3rd, Lee JY, Kang JD. Three-dimensional intervertebral kinematics in the healthy young adult cervical spine during dynamic functional loading. J Biomech 2015;48:1286-93.  Back to cited text no. 22
    
23.
Zhou C, Wang H, Wang C, Tsai TY, Yu Y, Ostergaard P, et al. Intervertebral range of motion characteristics of normal cervical spinal segments (C0-T1) during in vivo neck motions. J Biomech 2020;98:109418.  Back to cited text no. 23
    
24.
Swinkels RA, Swinkels-Meewisse IE. Normal values for cervical range of motion. spine (Phila Pa 1976) 2014;39:362-7.  Back to cited text no. 24
    
25.
Trott PH, Pearcy MJ, Ruston SA, Fulton I, Brien C. Three-dimensional analysis of active cervical motion: The effect of age and gender. Clin Biomech (Bristol, Avon) 1996;11:201-6.  Back to cited text no. 25
    
26.
Machino M, Yukawa Y, Imagama S, Ito K, Katayama Y, Matsumoto T, et al. Age-related and degenerative changes in the osseous anatomy, alignment, and range of motion of the cervical spine: A comparative study of radiographic data from 1016 patients with cervical spondylotic myelopathy and 1230 asymptomatic subjects. Spine (Phila Pa 1976) 2016;41:476-82.  Back to cited text no. 26
    
27.
Panjabi MM. The stabilizing system of the spine. Part II. Neutral zone and instability hypothesis. J Spinal Disord 1992;5:390-6.  Back to cited text no. 27
    
28.
Jacobs CA, Siepe CJ, Ito K. Viscoelastic cervical total disc replacement devices: Design concepts. Spine J 2020;20:1911-24.  Back to cited text no. 28
    
29.
Reyes-Sanchez A, Miramontes V, Olivarez LM, Aquirre AA, Quiroz AO, Zarate-Kalfopulos B. Initial clinical experience with a next-generation artificial disc for the treatment of symptomatic degenerative cervical radiculopathy. Sas J 2010;4:9-15.  Back to cited text no. 29
    
30.
Kim SH, Shin HC, Shin DA, Kim KN, Yoon DH. Early clinical experience with the Mobi-C disc prosthesis. Yonsei Med J 2007;48:457-64.  Back to cited text no. 30
    
31.
Staudt MD, Das K, Duggal N. Does design matter? Cervical disc replacements under review. Neurosurg Rev 2018;41:399-407.  Back to cited text no. 31
    
32.
Frost BA, Camarero-Espinosa S, Foster EJ. Materials for the spine: Anatomy, problems, and solutions. Materials (Basel) 2019;12:253.  Back to cited text no. 32
    
33.
Busscher I, Ploegmakers JJ, Verkerke GJ, Veldhuizen AG. Comparative anatomical dimensions of the complete human and porcine spine. Eur Spine J 2010;19:1104-14.  Back to cited text no. 33
    
34.
Dufour T, Beaurain J, Huppert J, Dam-Hieu P, Bernard P, Steib JP. Clinical and radiological evaluation of cervical disc arthroplasty with 5-year follow-up: A prospective study of 384 patients. Eur Spine J 2019;28:2371-9.  Back to cited text no. 34
    
35.
US Food and Drug Administration. M6-C Summary of Safety and Effectiveness Data. US Food and Drug Administration. Available from: https://www.accessdata.fda.gov/cdrh_docs/pdf17/P170036B.pdf. [Last accessed on 2020 Dec 26].  Back to cited text no. 35
    
36.
Hisey MS, Bae HW, Davis R, Gaede S, Hoffman G, Kim k, et al. Multi-center, prospective, randomized, controlled investigational device exemption clinical trial comparing Mobi-C cervical artificial disc to anterior discectomy and fusion in the treatment of symptomatic degenerative disc disease in the cervical spine. Int J Spine Surg 2014;8:7.  Back to cited text no. 36
    
37.
Davis RJ, Nunley PD, Kim KD, Hisey MS, Jackson RJ, Bae HW, et al. Two-level total disc replacement with Mobi-C cervical artificial disc versus anterior discectomy and fusion: A prospective, randomized, controlled multicenter clinical trial with 4-year follow-up results. J Neurosurg Spine 2015;22:15-25.  Back to cited text no. 37
    
38.
Huppert J, Beaurain J, Steib JP, Bernard P, Dufour T, Hovorka I, et al. Comparison between single- and multi-level patients: Clinical and radiological outcomes 2 years after cervical disc replacement. Eur Spine J 2011;20:1417-26.  Back to cited text no. 38
    


    Figures

  [Figure 1], [Figure 2], [Figure 3]
 
 
    Tables

  [Table 1], [Table 2], [Table 3], [Table 4]



 

Top
  
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    Abstract
   Introduction
    Materials and Me...
   Results
   Discussion
   Conclusion
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed1473    
    Printed36    
    Emailed0    
    PDF Downloaded126    
    Comments [Add]    

Recommend this journal